top of page

MinXRay HF100H+

Portable x-ray generator, up to 100kV output – AC POWERED

MinXRay HF100H+

The HF100H+ portable x-ray unit defines the industry standard for superior image quality, portability, and ease of use in any scenario. Easily programmed for storage of five different exposure techniques, the HF100H+ delivers state-of-the-art imaging and dependable performance in any setting. Compatible with all film and digital capture systems.


Features:

With a full-wave rectified high-frequency inverter, the HF100H+ is MinXray’s most cost-effective portable X-ray unit for medical imaging. The key feature of the HF100H+ is the output is independent of line voltage variations due to built-in line compensation. During circumstances where line power is low, you can be assured the X-ray will be captured and retain superior image quality.
• CPU control of x-ray tube
• 5 memory stations for storing X-ray exposure techniques
• Automatic, dynamic line compensation
• Compact, lightweight at 96.5 lbs. (44kg.) with the XGS MKIV LW portable stand
• Seamless integration with DR system is available


Product Specifications:
HF100H+
• Construction: Monobloc HF generator, full bridge inverter

• Output: 30 mA @ 40-60 kVDC,
25 mA @ 62-80 kVDC,
20 mA @ 82-100 kVDC

• kVDC Steps: 2KV intervals
• mAs: 0.6 - 120
• Timer: Ultra High Resolution 0.03-4.0 sec, 0.01 increments
• X-Ray Tube: Toshiba D-124S
• Focal Spot: 1.2 mm
• Anode Heat Storage: 20 kHU
• Total Filtration Incl. Collimator: 3.8 mm AI equivalent
• Inverter Frequency: 60 kHz High Frequency
• Line Voltage: AC Single Phase 100 - 140 V; 50/60 Hz
• Fuse Rating: 30 amp @ 125 kV
• Electrical Consumption: 20 amp @ 125 kV
• Maximum Consumption: 2.0 kW @ 100 kVDC, 20 mA
• Tubehead Dimensions: 11.5" x 9.1" x 17.7" (29.2 x 23 x 45 cm)
• Tubehead Weight: 44.5 lb. (20.2 kg)
• System Weight (HF100H+/XGS MKIV LW Stand): 86.5 lb (39.2 kg)
• Collimator: Continuously adjustable light beam type with central X-ray indicator
• Power Cord: 20 ft. (6.1 m)
• Exposure Switch: Two-stage, deadman with RJ11 connector
• Exposure Cord: 12 ft. (3.66 m)
• Compatibility: All film and digital capture systems